Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.

Recall: Z-1847-2014 · Initiated May 28, 2014

Recall

Recall Number
Z-1847-2014
Event Number
68256
Firm
Baxter Corporation Englewood
FEI Number
1419106
Product Code
NEP
Status
Terminated
Root Cause
Software design
Initiated
May 28, 2014
Posted
June 22, 2014
Terminated
August 8, 2016
Address
14445 Grasslands Dr, Englewood, CO, 80112-7062

Description

Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.

Reason

Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.

Action

Baxter sent an Urgent Device Correction letter dated May 27, 2014 via certified mail to affected customers. The letter identified the affected product, problem and actios to be taken. Customers were instructed to: 1. Review and distribute this communication to pharmacy staff and compounding staff that use ABACUS software. This notice is intended only for customers that have upgraded to version 3.1 of the software. 2. Please acknowledge receipt of this notification by completing the enclosed customer reply form and return it to Baxter by either faxing it to 1-224-270-5457 or scanning and e-mailing it to [email protected]. 3. If you have distributed ABACUS to other facilities, please notify the end users in accordance with your customary procedures. For questions call Baxter Techenical Support line at 1-800-678-2292.

Distribution

US Nationwide Distribution and the state of VA.

Quantity

42