ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position, and to provide a beam hold signal in case of a deviations in order to treat lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Recall: Z-1846-2021 · Initiated May 14, 2021

Recall

Recall Number
Z-1846-2021
Event Number
87984
Firm
Brainlab AG Olof-palme-str. 9 Munich Germany
FEI Number
3002619595
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
May 14, 2021
Terminated
March 20, 2023

Description

ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position, and to provide a beam hold signal in case of a deviations in order to treat lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Reason

Display of potential patient movement might be delayed to the user for high dose treatments.

Action

An Urgent Field Safety Notice/Product Notification dated 4/29/2021 was issued on 5/14/2021 via email explaining the safety issue and letting the customer know that this issue is in addition to the previous notification of 9/8/2020 which has now expanded to versions 1.0.1 and 1.0.2.

Distribution

US Nationwide distribution in the states of AZ, AR, CA, FL, IN, NJ, NY, NC, OR, PA, and TX. There was no government/military distribution.

Quantity

22 systems in the U.S.