FDA Recall Terminated

VersaTREK Instrument Series 528 REF 6528, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.

Recall: Z-1842-2015 · Initiated May 19, 2015

Recall

Recall Number
Z-1842-2015
Event Number
71291
Firm
Remel Inc
FEI Number
1924669
Product Code
MDB
Status
Terminated
Root Cause
Release of Material/Component prior to receiving test results
Initiated
May 19, 2015
Posted
June 25, 2015
Terminated
March 24, 2017
Address
12076 Santa Fe Trail Dr, Lenexa, KS, 66215-3519

Description

VersaTREK Instrument Series 528 REF 6528, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.

Reason

Use of the recalled product may result in false positive reports.

Action

Remel, Inc. a part of Thermo Scientific sent a Product Correction Notification dated May 19, 2015, to all affected customers. The notification states the problem with the device and that the recalling firm's Technical Services department will be contacting the customer to schedule a field service of the device to correct the issue. Customers were instructed to notify any personnel who need to be notified of the potential for a false positive resulting from rare inconsistent barometric pressure readings. Return the attached Medical Device Recall Return Response form acknowledging receipt of the notice of the affected product. If product was further distributed customers were instructed to contact those entities, advise them of the situation and provide them with a copy of the letter. Customers with questions were instructed to contact the Technical Services Department at 800-642-7029. For questions regarding this recall call 770-409-0713.

Distribution

Worldwide distribution: US (nationwide) to states of AZ, CO, FL, GA, IA, ID, LA, MI, MO, PR, TN, TX, VA, and WA and the countries of Canada, People's Republic of China, Egypt, Finland, France, Indonesia, Israel, Jordan, Kenya, Republic of Korea, Latvia, Romania, Russia, Spain, Thailand, United Kingdom, Uruguay, and Vietnam.

Quantity

21 devices