Siemens LANTIS System ; LANTIS Commander, 10 Users ; LANTIS 6.1 Commander ; LANTIS 8.3 Commander The intended use of the linear accelerator is to deliver x-ray radiation for therapeutic treatment of cancer.
Recall
- Recall Number
- Z-1840-2014
- Event Number
- 68379
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Use error
- Initiated
- May 6, 2014
- Posted
- June 20, 2014
- Terminated
- December 15, 2014
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
Siemens LANTIS System ; LANTIS Commander, 10 Users ; LANTIS 6.1 Commander ; LANTIS 8.3 Commander The intended use of the linear accelerator is to deliver x-ray radiation for therapeutic treatment of cancer.
There is a potential risk when using LANTIS OIS System client software with operating systems for which it has not been validated and released, resulting in incorrectly stored data, such as treatment records, and LANTIS database corruption.
Siemens sent an Urgent Medical Device Correction Field Safety Notice dated was sent to end users on/about May 6, 2014, to notify them of the safety issue and provide instructions to avoid potential risks. Customers were instructed to make sure all computers with the LANTIS client software installed are running on an operating system that was validated by Siemens for opration with LANTIS, e.g. Microsoft Windows XP Service Pack 3 (SP3). Customers were asked to include the Field Safety Notice in their System Owner Manual chapter "Safety Advisory Letters" where it should remain and that all affected personnel affected be informed. For questions regarding this recall call 1-610-219-6300.
Nationwide Distribution
108