FDA Recall Terminated

Siemens LANTIS System ; LANTIS Commander, 10 Users ; LANTIS 6.1 Commander ; LANTIS 8.3 Commander The intended use of the linear accelerator is to deliver x-ray radiation for therapeutic treatment of cancer.

Recall: Z-1840-2014 · Initiated May 6, 2014

Recall

Recall Number
Z-1840-2014
Event Number
68379
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IYE
Status
Terminated
Root Cause
Use error
Initiated
May 6, 2014
Posted
June 20, 2014
Terminated
December 15, 2014
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Siemens LANTIS System ; LANTIS Commander, 10 Users ; LANTIS 6.1 Commander ; LANTIS 8.3 Commander The intended use of the linear accelerator is to deliver x-ray radiation for therapeutic treatment of cancer.

Reason

There is a potential risk when using LANTIS OIS System client software with operating systems for which it has not been validated and released, resulting in incorrectly stored data, such as treatment records, and LANTIS database corruption.

Action

Siemens sent an Urgent Medical Device Correction Field Safety Notice dated was sent to end users on/about May 6, 2014, to notify them of the safety issue and provide instructions to avoid potential risks. Customers were instructed to make sure all computers with the LANTIS client software installed are running on an operating system that was validated by Siemens for opration with LANTIS, e.g. Microsoft Windows XP Service Pack 3 (SP3). Customers were asked to include the Field Safety Notice in their System Owner Manual chapter "Safety Advisory Letters" where it should remain and that all affected personnel affected be informed. For questions regarding this recall call 1-610-219-6300.

Distribution

Nationwide Distribution

Quantity

108