FDA Recall Open, Classified

Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170085

Recall: Z-1834-2023 · Initiated May 25, 2023

Recall

Recall Number
Z-1834-2023
Event Number
92373
Firm
TELEFLEX LLC
FEI Number
3005747797
Product Code
BTR
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 25, 2023
Posted
June 29, 2023
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437

Description

Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170085

Reason

Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.

Action

Teleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 05/25/2012 by FedEx. The notice explained the problem, the risk, and requested the following: Users should cease use and distribution of affected product and immediately quarantine the affected product. Distributors were directed to cease use, distribution, and quarantine the product, and to notify their customers. Teleflex is seeking the return of the affected product. For further information: Contact: Customer Service, Telephone: 1-866-396-2111 FAX: 1-855-419-8507, Email: [email protected]

Distribution

US Nationwide distribution including Puerto Rico.

Quantity

4370 units