FDA Recall Open, Classified

CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.

Recall: Z-1831-2024 · Initiated April 15, 2024

Recall

Recall Number
Z-1831-2024
Event Number
94450
Firm
Maquet Medical Systems USA
FEI Number
3008355164
Product Code
DTQ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 15, 2024
Posted
May 16, 2024
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.

Reason

The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.

Action

Consignees received an "URGENT MEDICAL DEVICE - Removal" email, dated 4/15/2024, notifying them of this recall event. Consignees are asked to examine their inventory to identify if they possess any affected devices and all consignees are asked to return the provided Response Form via email to identify where affected devices are located. Consignees that report having any affected devices will be contacted by Getinge to arrange for return of the affected device to upgrade the system's sensor panel and perform testing to ensure that leakage current is within specifications. Technical Support can also be contacted at 888-943-8872 (option 4, option 2, option 1) Monday through Friday between 8:00 AM and 6:00 PM EST to make service arrangements. Consignees are asked to not use affected devices for any additional patients until it has been cleared by the MCP/Getinge system check, repaired, or replaced. The provided recall notice is to be forwarded to users of the CARDIOHELP-i system and if the device was further distributed.

Distribution

US Nationwide distribution in the states of NJ & TX.

Quantity

148 units (US: 5; OUS: 143)