LapSac Surgical Tissue Pouch. The current intended use for the LapSac Surgical Tissue Pouch is for isolating tissue during, or prior to, surgical removal and/or morcellation. The device is a reinforced pouch with an inner liner and drawstring closure. The device may be used to isolate tissue and/or act as a containment system during surgery. It is provided in a variety of sizes and has applications in the Urology, Surgery, and Reproductive Health business units.
Recall
- Recall Number
- Z-1830-2015
- Event Number
- 71287
- Firm
- Cook Inc.
- FEI Number
- 1820334
- Product Code
- KGY
- Status
- Terminated
- Root Cause
- Labeling Change Control
- Initiated
- May 14, 2015
- Posted
- June 24, 2015
- Terminated
- September 12, 2016
- Address
- 750 N Daniels Way, Bloomington, IN, 47404-9120
Description
LapSac Surgical Tissue Pouch. The current intended use for the LapSac Surgical Tissue Pouch is for isolating tissue during, or prior to, surgical removal and/or morcellation. The device is a reinforced pouch with an inner liner and drawstring closure. The device may be used to isolate tissue and/or act as a containment system during surgery. It is provided in a variety of sizes and has applications in the Urology, Surgery, and Reproductive Health business units.
Includes Information not stated in the intended use.
Cook Medical sent an "Recall Communication Updated Instructions for Use" letter dated May 15, 2015, to all affected customers . The letter identified the product the problem and the action needed to be taken by the customer. Cook has requested that customers do the following: The intended use for the device is being modified from "Used to isolate tissue" during, or prior to, surgical removal and/or morcellation to "Used to isolate tissue" during, or prior to, surgical removal. A contraindication being added is as follows: " Not for use with power morcellators. The warnings are being modified to read as follows: " The LapSac Surgical Tissue Pouch is not intended for use with sharp surgical instruments. " Care should be taken to avoid damaging, puncturing or tearing the LapSac Surgical Tissue Pouch. " Use caution when removing potentially malignant or harmful tissues, to isolate contamination from surrounding tissue. The instructions for use are being modified to read as follows: 1. Following aseptic technique, introduce the LapSac Surgical Tissue Pouch into the abdominal cavity. 2. Place tissue to be removed from the body into the pouch. 3. Using the drawstring, carefully close the LapSac Surgical Tissue Pouch. Please note that if excess force is used, the drawstring may break or tear the LapSac. 4. Remove the LapSac Surgical Tissue Pouch from the body, taking care not to spill contents of the bag. Please report any Adverse Event to Cook Medical Customer Relations (800) 457-4500 or 1-812-339-2235. Monday through Friday between 7:30 a.m. and 5:00 p.m. Eastern Daylight Time or email at [email protected]; and to FDA. The customer notice must be passed on to appropriate personnel within each organization or to any organization where the potentially affected devices have been transferred. If customers have further questions regarding labelling changes for this device, please contact Cook Sales Representative or Cook M
Worldwide Distribution - US (nationwide) including PR and to the states of: AZ, AK, AL, AR, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, MO, ME, MT, NC, ND, NE, NH, NJ, NY, NM, OH,OK,OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY., and to the countries of : China, Korea, Denmark, Mexico, New Zealand, India and Colombia.
123,184 devices were distributed in commerce