EGIA 60 ARTICULATING MED THICK SULU , EGIA60AMT
Recall
- Recall Number
- Z-1829-2019
- Event Number
- 82950
- Firm
- Covidien LLC
- FEI Number
- 1219930
- Product Code
- GDW
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- May 17, 2019
- Address
- 60 Middletown Ave, North Haven, CT, 06473-3908
Description
EGIA 60 ARTICULATING MED THICK SULU , EGIA60AMT
The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
All consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Endo GIA" Articulating Reloads with Tri-Staple Technology. " Complete the Recalled Product Return Form. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Nationwide domestic distribution, worldwide foreign distribution.
3,113,280