FDA Recall Terminated

Interventional Pain 6" 13 G (2.4 mm) Percutaneous Discectomy Probe, REF 407-265, Sterile, Stryker Instruments, Kalamazoo, MI. The DeKompressor Percutaneous Discectomy Probe is intended for use in the apiration of disc material during percutaneous discectomies in the lumbar, thoracic, and cervical regions of the spine

Recall: Z-1829-2011 · Initiated February 3, 2011

Recall

Recall Number
Z-1829-2011
Event Number
57899
Firm
Stryker Instruments Division of Stryker Corporation
FEI Number
1811755
Product Code
HRX
Status
Terminated
Root Cause
Other
Initiated
February 3, 2011
Posted
March 28, 2011
Terminated
June 1, 2011
Address
4100 East Milham Ave., Kalamazoo, MI, 49001

Description

Interventional Pain 6" 13 G (2.4 mm) Percutaneous Discectomy Probe, REF 407-265, Sterile, Stryker Instruments, Kalamazoo, MI. The DeKompressor Percutaneous Discectomy Probe is intended for use in the apiration of disc material during percutaneous discectomies in the lumbar, thoracic, and cervical regions of the spine

Reason

The firm received five complaints that the stylet did not fit down the cannula. The investigation revealed that the incorrect probe assembly was packaged for two DeKompressor part numbers with the lot number of 10214012.

Action

Stryker Instruments Division of Stryker Corporation sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letter dated February 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately check their inventory for any recalled product and quarantine any affected product. If the product was further distributed, customers were asked to forward the letter and the completed Reply Form to all affected locations. Customers were to be contacted to coordinate return of the affected product. Upon receipt of the product, a replacement would be issued to their account. For questions regarding this recall call 1-800-253-3210.

Distribution

Worldwide Distribution - USA including the states of PA, NY, KS, NJ, TX, MI, OK, and FL, and the countries of Italy and Switzerland.

Quantity

44 units both products