FDA Recall Open, Classified

CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.

Recall: Z-1828-2026 · Initiated March 6, 2026

Recall

Recall Number
Z-1828-2026
Event Number
98476
Firm
Integra LifeSciences Corp. (NeuroSciences)
FEI Number
3003418325
Product Code
LFL
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 6, 2026
Posted
April 17, 2026
Address
1100 Campus Rd, Princeton, NJ, 08540-6650

Description

CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.

Reason

Software issue that renders the touch screen unresponsive.

Action

An URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice dated 3/6/26 was sent to consignees. The notice instructs consignees to restart systems when freeze to restore functionality of the touchscreen capabilities for 8 minutes; this action may be used multiple times. Integra's Service and Repair can provide remote support to disable and re-enable the write filter with Integra employees and/or Authorized service personnel. This will reset the console to provide approximately 28 days with no touchscreen issue. Integra Lifesciences Service Team will contact consignees within 5 to 10 business days to schedule correction of affected devices at no cost. Consignees with any questions can direct them to Customer Service by emailing [email protected] or by phone at 1-800-654-2873.

Distribution

Worldwide - US Nationwide distribution in the states of MI & NY. The countries of Belgium, China, Czechia, Dubai, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands. Poland, Spain, Switzerland, Tunisia & United Kingdom.

Quantity

78 units