FDA Recall Terminated

MIM 4.0 and MIM 4.1 is a software package designed for use in diagnostic imaging. It is a stand alone software package which operates on Windows 2000/XP.

Recall: Z-1827-2008 · Initiated April 23, 2008

Recall

Recall Number
Z-1827-2008
Event Number
48114
Firm
MIMvista Corp
FEI Number
3004363352
Product Code
LLZ
Status
Terminated
Root Cause
Software change control
Initiated
April 23, 2008
Posted
September 3, 2008
Terminated
April 14, 2015
Address
25200 Chagrin Blvd, Ste 200, Cleveland, OH, 44122-5684

Description

MIM 4.0 and MIM 4.1 is a software package designed for use in diagnostic imaging. It is a stand alone software package which operates on Windows 2000/XP.

Reason

In MIM (4.0.2-4.0.5) and MIM (4.1.0-4.1.2) versions, there is an error that may occur during the calculation of statistics using the SUV Tool or Contour Statistics. It will occur only under the following specific circumstance: A fusion of a primary volume which has a Z-Axis slice separation larger than that of the secondary volume and statistics are reported on the secondary volume.

Action

MIMvista Corp has issued this safety recall notification for MIM versions 4.0.2 through 4.1.2. An error has been discovered in the way SW and contour statistics are measured and reported under some specific circumstances. Users of these versions should immediately contact MIMvista support at (866) 421-2536 or visit httv://www.mimvista.com/recall/ to obtain complete instructions regarding this error. A new version that corrects this problem is available on the Downloads page, http://www,mimvista.com/download

Distribution

The product was sold in the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI & WV. It was also sold in Europe, Asia and Canada.

Quantity

221