FDA Recall Open, Classified

'n'-Hole Post which is a component of the Revolution External Fixation System Catalog Numbers: 100801, 100802, 100803, 100804, 100805

Recall: Z-1824-2023 · Initiated July 23, 2019

Recall

Recall Number
Z-1824-2023
Event Number
92390
Firm
New Standard Device Inc
FEI Number
3015536247
Product Code
KTT
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 23, 2019
Posted
June 9, 2023
Address
4848 Research Dr, San Antonio, TX, 78240-5005

Description

'n'-Hole Post which is a component of the Revolution External Fixation System Catalog Numbers: 100801, 100802, 100803, 100804, 100805

Reason

The threads on posts stripping out when tensioned during application process.

Action

On 07/23/2019, the firm emailed a "FIELD SAFETY NOTICE" Letter to customers informing them that threads on posts were stripping out when tensioned during application process. Customers were instructed to: 1. Remove remaining POSTS from their inventory and return to Metalogix Inc. at 4848 Research Drive, San Antonio, TX 78240 Please label return package "ATTENTION: QA Dept" Metalogix Fed-Ex Account# 845 986 983 Metalogix will expedite new product back to customers For further information or questions, contact: Rachel Grimes, Quality Assurance Manager at 833-659-2019, EXT 6 [email protected]

Distribution

U.S. Nationwide distribution in the states of AZ, CT, OH, TX, VA.

Quantity

222 units