'n'-Hole Post which is a component of the Revolution External Fixation System Catalog Numbers: 100801, 100802, 100803, 100804, 100805
Recall
- Recall Number
- Z-1824-2023
- Event Number
- 92390
- Firm
- New Standard Device Inc
- FEI Number
- 3015536247
- Product Code
- KTT
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- July 23, 2019
- Posted
- June 9, 2023
- Address
- 4848 Research Dr, San Antonio, TX, 78240-5005
Description
'n'-Hole Post which is a component of the Revolution External Fixation System Catalog Numbers: 100801, 100802, 100803, 100804, 100805
The threads on posts stripping out when tensioned during application process.
On 07/23/2019, the firm emailed a "FIELD SAFETY NOTICE" Letter to customers informing them that threads on posts were stripping out when tensioned during application process. Customers were instructed to: 1. Remove remaining POSTS from their inventory and return to Metalogix Inc. at 4848 Research Drive, San Antonio, TX 78240 Please label return package "ATTENTION: QA Dept" Metalogix Fed-Ex Account# 845 986 983 Metalogix will expedite new product back to customers For further information or questions, contact: Rachel Grimes, Quality Assurance Manager at 833-659-2019, EXT 6 [email protected]
U.S. Nationwide distribution in the states of AZ, CT, OH, TX, VA.
222 units