FDA Recall Terminated

Headset Model Number RTC-551-070 for use with the Serene Sound Digital MRI Compatible High Fidelity Stereo Sound System .

Recall: Z-1824-2008 · Initiated May 14, 2007

Recall

Recall Number
Z-1824-2008
Event Number
47834
Firm
Resonance Technology Inc
FEI Number
1000306474
Product Code
LNH
Status
Terminated
Root Cause
Device Design
Initiated
May 14, 2007
Posted
September 2, 2008
Terminated
July 7, 2011
Address
18121 Parthenia St, Northridge, CA, 91325-3350

Description

Headset Model Number RTC-551-070 for use with the Serene Sound Digital MRI Compatible High Fidelity Stereo Sound System .

Reason

Potential safety concerns - Use of this product with a MRI may result in the headset's cord overheating and patients experiencing burns to the skin.

Action

The strategy was notifying customers by written communication. The letter advises Resonance Technology, Inc. is voluntarily recalling the MRI Headset, Model No. RTC-2K and RTC- 551-070 because of potential safety concerns. Use of this product with a MRI may result in the headset's cord overheating and patients experiencing burns to skin. The letter also advises that Resonance Technology Inc. has designed and developed the significantly improved MRI Headset, Model No. RTC-661-060-09P-001 (upgrade) and began shipping this product in January 2007. Customers who received the Headset Model; RTC-2K and RTC- 551-070, are asked to upgrade to MRI Headset Model No. RTC-661-060-09P-001. Response is requested as to the number of headsets to be replaced. In addition, consignees are advised that if they further distributed this product, to identify their customers and notify them at once of this product recall. If you have questions, contact Olivia Shayesteh at 818-882-1997 or via email at [email protected]

Distribution

Nationwide and Internationally: Canada, Japan, Taiwan, United Kingdom, Netherlands, Italy, Spain, Germany, Switzerland, France,

Quantity

29 units