Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Recall
- Recall Number
- Z-1823-2026
- Event Number
- 98619
- Firm
- Gentuity, LLC
- FEI Number
- 3011578544
- Product Code
- DQO
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- March 23, 2026
- Posted
- April 16, 2026
- Address
- 142 North Rd, Ste G, Sudbury, MA, 01776-1142
Description
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
On March 23, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: The customers will be required to follow the updated instructions to correct for repeated frames per the revised Operator Manual to address longitudinal measurement accuracy until the updated version of the software that addresses this issue is installed on the console within 6 months.
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX and the countries of Belgium, Japan.
184 units (28 US, 156 OUS)