FDA Recall Terminated

Tapered SwissPlus Implant System, Model No. SPB10

Recall: Z-1823-2017 · Initiated March 20, 2017

Recall

Recall Number
Z-1823-2017
Event Number
76746
Firm
Zimmer Dental Inc
FEI Number
2023141
Product Code
DZE
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 20, 2017
Terminated
November 1, 2017
Address
1900 Aston Ave, Carlsbad, CA, 92008-7308

Description

Tapered SwissPlus Implant System, Model No. SPB10

Reason

It was manufactured with the minor diameter of the internal thread that is below specification.

Action

A customer letter was sent on 3/20/17 to inform customers that Zimmer Biomet is recalling a single lot of the Tapered SwissPlus 3.7mmD Implant, Lot No. 63498673. The letter informs customers that the recall is due to one (1) lot of part number SPB10 (lot 63498673) that was manufactured with the minor diameter of the internal thread below specification. This condition may cause mating issues with some screws. The condition occurs when the screw used has a minor thread diameter that is closer to the maximum allowed diameter (maximum material condition). However, screws closer to the least material condition will still fit into the implant. Customers with questions or concerns are instructed to call the customer call center at 1-800-854-7091 between 8:00 am and 5:00pm EST, Monday through Friday. Customers are also instructed to email to [email protected].

Distribution

Worldwide: ZA, FR, IT, IR

Quantity

188 units