The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031
Recall
- Recall Number
- Z-1822-2017
- Event Number
- 76810
- Firm
- Philips Electronics North America Corporation
- FEI Number
- 1218950
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- March 22, 2017
- Terminated
- May 17, 2021
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032
Description
The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Consignees were notified of the recall via letter sent on 3/22/2017. The letter explains the issue, actions that should be taken by the customer, and actions to be taken by Philips.
Nationwide, worldwide including Canada
6992 (total distributed for all products)