UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039
Recall
- Recall Number
- Z-1821-2017
- Event Number
- 76810
- Firm
- Philips Electronics North America Corporation
- FEI Number
- 1218950
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- March 22, 2017
- Terminated
- May 17, 2021
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032
Description
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Consignees were notified of the recall via letter sent on 3/22/2017. The letter explains the issue, actions that should be taken by the customer, and actions to be taken by Philips.
Nationwide, worldwide including Canada
6992 (total for all devices)