FDA Recall Terminated

UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039

Recall: Z-1821-2017 · Initiated March 22, 2017

Recall

Recall Number
Z-1821-2017
Event Number
76810
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
OWB
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
March 22, 2017
Terminated
May 17, 2021
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039

Reason

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Action

Consignees were notified of the recall via letter sent on 3/22/2017. The letter explains the issue, actions that should be taken by the customer, and actions to be taken by Philips.

Distribution

Nationwide, worldwide including Canada

Quantity

6992 (total for all devices)