FDA Recall Terminated

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

Recall: Z-1821-2012 · Initiated May 11, 2012

Recall

Recall Number
Z-1821-2012
Event Number
61877
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
IXR
Status
Terminated
Root Cause
Device Design
Initiated
May 11, 2012
Posted
June 18, 2012
Terminated
October 6, 2016
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

Reason

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

Action

Philips sent an Urgent - Field Safety Notice letter dated May 11, 2012 to their affected customers. The letter identified the affected product, what the issues are, under what circumstances they may occur and the actions to be taken to avoid or minimize the occurrence of the issue Philips plans to supply an Addendum to the Instruction for Use. A Philips Service Engineer will contact the affected customers as soon as the Field Action Kit is ready to be implemented. For questions contact your local Philips representative: Technical Support Line 1-800-722-9377.

Distribution

6 US customers

Quantity

6