FDA Recall Open, Classified

VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820

Recall: Z-1819-2024 · Initiated April 9, 2024

Recall

Recall Number
Z-1819-2024
Event Number
94436
Firm
QUIDELORTHO
FEI Number
3027243135
Product Code
JIS
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 9, 2024
Posted
May 14, 2024
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820

Reason

Free T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.

Action

QuidelOrtho issued Important Product Correction Notification issued April 9, 2024. Letter states reason for recall, health risk and action to take: Discontinue using, render unusable, and discard affected lots of VITROS Free T3 Reagent Pack (and associated calibrators). " Complete the enclosed Confirmation of Receipt form no later than April 17, 2024. Upon receipt of your completed form, QuidelOrtho will provide credit for, or replacement of, discarded product. " Save this notification with your User Documentation or post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System until the issue has been resolved. " Please forward this notification if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Resolution QuidelOrtho s investigation is on-going, and currently working to identify root cause. Contact Global Services Organization at 1-800-421-3311.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore , Spain, Sweden, The Netherlands, United Kingdom.