FDA Recall Terminated

Q-Tel. The Q-Tel RMS system is a computer-based cardiac and pulmonary rehabilitation data acquisition and editing system.

Recall: Z-1818-2020 · Initiated March 18, 2020

Recall

Recall Number
Z-1818-2020
Event Number
85297
Firm
WELCH ALLYN, INC/MORTARA
FEI Number
2183461
Product Code
DPS
Status
Terminated
Root Cause
Device Design
Initiated
March 18, 2020
Posted
April 29, 2020
Terminated
February 9, 2026
Address
7865 N 86th St, Milwaukee, WI, 53224-3431

Description

Q-Tel. The Q-Tel RMS system is a computer-based cardiac and pulmonary rehabilitation data acquisition and editing system.

Reason

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Action

Welch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed.

Distribution

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Quantity

78,521 units total