FDA Recall Open, Classified

ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.

Recall: Z-1817-2025 · Initiated December 22, 2023

Recall

Recall Number
Z-1817-2025
Event Number
96781
Firm
Onkos Surgical, Inc.
FEI Number
3013450937
Product Code
JDI
Status
Open, Classified
Root Cause
Labeling Change Control
Initiated
December 22, 2023
Posted
May 23, 2025
Address
77 E Halsey Rd, Parsippany, NJ, 07054-3714

Description

ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.

Reason

Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.

Action

ONKOS Surgical notified their distributors on about 12/22/2023 via email. They were instructed to segregate and quarantine the product immediately, complete and return the "Product and Distribution Information" form within 2 business days and return any existing implants on hand within 2 business days after signing the acknowledgement. They were also instructed that they will receive replacement implants and to place them in their modular segmental collar stem implant packs.

Distribution

US Nationwide distribution in the states of California, Kentucky, and Wisconsin.

Quantity

23 units