FDA Recall Terminated

BD Viper LT System, catalog number 442839 when used in conjunction with the BD ProbeTec Chlamydia trachomatis (CT) Qx Assay (cat. #442959) and the BD ProbeTec Neisseria gonorrhoeae (GC) Qx Assay (cat. #442842). In vitro diagnostic.

Recall: Z-1816-2015 · Initiated November 7, 2014

Recall

Recall Number
Z-1816-2015
Event Number
70164
Firm
Becton Dickinson & Co. BD Diagnostic Systems
FEI Number
1119779
Product Code
OOI
Status
Terminated
Root Cause
Device Design
Initiated
November 7, 2014
Posted
June 18, 2015
Terminated
May 9, 2016
Address
7 Loveton Circle, Sparks, MD, 21152-0999

Description

BD Viper LT System, catalog number 442839 when used in conjunction with the BD ProbeTec Chlamydia trachomatis (CT) Qx Assay (cat. #442959) and the BD ProbeTec Neisseria gonorrhoeae (GC) Qx Assay (cat. #442842). In vitro diagnostic.

Reason

Potential for false positive Chlamydia trachomatis (CT) results and Neisseria gonorrhoeae (GC) results when processing the CT Qx and GC Qx Assays on the BD Viper LT Instrument.

Action

BD sent an Urgent Field Corrective Action letter dated November 2014, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer. Customers were advised: a. Include the date recall was initiated, if it is already underway or the date your firm plans to start the recalling process. b. How do you plan to notify all the consignees affected by this recall? (press release, letter, telefax, telephone, e-mail, visit, etc.) The customer was initially contacted via a phone call and subsequently provided with a written communication. c. How do you plan to monitor the number of consignees non-responding to the recall communication? The single US customer was contacted by phone and provided the written communication. d. How do you plan to do effectiveness checks of this recall? The affected instrument will be corrected. e. Date your firm ceased further distribution of the product(s). Product was not placed on hold. This issue does not affect new customers. f. How do you plan to store the recall product? BD has not requested that instruments or reagents be returned. g. How do you plan to dispose of the recall products? BD is not requesting that customers return the instruments or reagents. If you require further assistance or clarification, please contact the BD Technical Services Department at 1-800 638-8663.

Distribution

Worldwide distribution to US (state of LA), Hong Kong, Austria, China, France, and the UK.

Quantity

7 Viper LT Systems