FDA Recall Open, Classified

Autosoft 90, model numbers 1000282, 1002817, 1002818, 1002819, 1002820, 1002821, 1002822, 1002823, and 1002824. Infusion sets used with Tandem Insulin Subcutaneous Infusion Pump.

Recall: Z-1810-2022 · Initiated July 21, 2022

Recall

Recall Number
Z-1810-2022
Event Number
90719
Firm
Unomedical A/S Aaholmvej
FEI Number
1000477110
Product Code
FPA
Status
Open, Classified
Root Cause
Labeling Change Control
Initiated
July 21, 2022
Posted
September 26, 2022
Address
3 OSTED LEJRE Denmark, 1

Description

Autosoft 90, model numbers 1000282, 1002817, 1002818, 1002819, 1002820, 1002821, 1002822, 1002823, and 1002824. Infusion sets used with Tandem Insulin Subcutaneous Infusion Pump.

Reason

The IFU incorrectly states the insulin fill value of two sizes of cannula which can result in an initial insulin infusion to be fewer units than intended.

Action

The recalling firm issued letters dated 21 July 2022 on the same day via email. The letter was to the attention of distributors and end users/consumers, but the recalling firm issued the letter only to the distributor. The letter provided a description of the problem, risk to health, and actions to be taken. The actions listed were to immediately examine their inventory and quarantine the product lots listed in the attached list. The enclosed response form was to be completed and returned by email no later than 28-Jul-2022. If the product has been further distributed, the distributor is to identify their customers and notify them of the product correction using the patient letter provided. The distributor will also be sending health care providers a Unomedical notification to make the HCP aware that all patients who may be using the Autosoft 90 infusion set models will be notified of the error in the IFU.

Distribution

Distribution was made to California. There was no government/military distribution.

Quantity

1,643,320 units