FDA Recall Terminated

Medtronic Preimplantation Test Kit, Catalog No. 21047. The kit is indicated for use with Medtronic CSF-Flow Control Valves and Delta Valves, when a simple preimplantation test is desired to verify that the valve conforms to labeled product specifications.

Recall: Z-1810-2013 · Initiated June 28, 2013

Recall

Recall Number
Z-1810-2013
Event Number
65600
Firm
Medtronic Neurosurgery
FEI Number
3015531529
Product Code
KYZ
Status
Terminated
Root Cause
Incorrect or no expiration date
Initiated
June 28, 2013
Posted
July 25, 2013
Terminated
September 23, 2013
Address
125 Cremona Dr, Goleta, CA, 93117-3083

Description

Medtronic Preimplantation Test Kit, Catalog No. 21047. The kit is indicated for use with Medtronic CSF-Flow Control Valves and Delta Valves, when a simple preimplantation test is desired to verify that the valve conforms to labeled product specifications.

Reason

Medtronic Neurosurgery is recalling the Medtronic Preimplantation Test Kit because the outer carton had expiration dates that were incorrect. The date on the outer carton is later than the expiration date for one or more of the individually packaged sterile components in the kit.

Action

An "Urgent Field Safety Notice" recall letter dated 6/28/13 was sent to customers who purchased the Medtronic Preimplantation Test Kit, catalog number 21047, to inform them of the recall due to the incorrect outer carton expiration dates. The recall letter informed the customers of the problems and actions identified. Customers were instructed to contact the Quality Department at (805) 571-8725 for questions.

Distribution

Nationwide Distribution in the US.

Quantity

915 units