FDA Recall Terminated

Ulnar Shorting Saw Blade is labeled in part: "Congruent Bone Plates, SIZE Ulnar Shortening Saw Blade - Hub Style S, REF 80-0499-S, Material: Stainless Steel, STERILE, Rx Only, ACUMED, Hillsboro, OR, USA". The Ulnar Saw Blade is an accessory used with a Congruent Bone Plate System covered under K012655, but it is labeled and sold separately by itself. The saw blade is also sold as part of the Upper Extremity Osteotomy System (BOM 15-0030). The BOM 15-0030 system does not have a label, but the saw blade is individually labeled and shipped together with this system when it is ordered. Intended use: Cutting Bones

Recall: Z-1807-2010 · Initiated April 14, 2010

Recall

Recall Number
Z-1807-2010
Event Number
55410
Firm
Acumed LLC
FEI Number
1000125930
Product Code
HRS
Status
Terminated
Root Cause
Other
Initiated
April 14, 2010
Posted
June 16, 2010
Terminated
June 28, 2010
Address
5885 NW Cornelius Pass Rd, Hillsboro, OR, 97124-9432

Description

Ulnar Shorting Saw Blade is labeled in part: "Congruent Bone Plates, SIZE Ulnar Shortening Saw Blade - Hub Style S, REF 80-0499-S, Material: Stainless Steel, STERILE, Rx Only, ACUMED, Hillsboro, OR, USA". The Ulnar Saw Blade is an accessory used with a Congruent Bone Plate System covered under K012655, but it is labeled and sold separately by itself. The saw blade is also sold as part of the Upper Extremity Osteotomy System (BOM 15-0030). The BOM 15-0030 system does not have a label, but the saw blade is individually labeled and shipped together with this system when it is ordered. Intended use: Cutting Bones

Reason

The saw blades may not cut aggressively enough for their intended application therefore creating a risk that the saw could generate excessive heat, which may contribute to bone necrosis and/or non-union.

Action

On April 14, 2010, Acumed began contacting consignees by telephone. On April 15, 2010, Acumed sent out the URGENT NOTICE: DEVICE RECALL Initial Notification letter and Recalled Product Status Form to their consignees. The letter describes the recall product as Ulnar Shorting Saw Blade-Hub Style S, Part no. 80-0499-S and Lot Codes. The Initial Notification letter instructs consignees to return the product back to Acumed immediately and notify their sub-accounts (customers) if products have been further distributed. The consignees also are asked to fill out and send back the Recalled Product Status Form . Customers should call the Customer Service at 1-888-627-9957 or e-mail to [email protected] for any questions regarding the recall.

Distribution

Worldwide distribution: USA and to the following countries: Spain, Great Britain, South Africa, Puerto Rico, Canada, Turkey, Ireland, Finland, Belgium, Australia, Germany, and Sweden.

Quantity

718 units