RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Recall: Z-1806-2014 · Initiated April 25, 2014

Recall

Recall Number
Z-1806-2014
Event Number
68162
Firm
RAYSEARCH LABORATORIES AB Sveavaegen 9 Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
April 25, 2014
Posted
June 17, 2014
Terminated
June 26, 2017

Description

RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Reason

This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.

Action

On April 25, 2014, RaySearch emailed all consignees of the affected product a copy of the recall notice as well as the response forms. Back up notification methods include phone calls and site visits.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.

Quantity

25 sites/245 units (US), 25 sites/124 units (foreign)