BD BBL Anaerobe CNA Agar with 5% Sheep Blood // Anaerobe Laked Sheep Blood KV Agar, packaged in cartons of 20 plates, Catalog # 297041, and labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 USA 800-638-8663 www.bd.com/ds*** Anaerobe Laked Sheep Blood KV Agar is used in a qualitative procedure for isolation of gram-negative anaerobic bacilli.
Recall
- Recall Number
- Z-1801-2013
- Event Number
- 65614
- Firm
- Becton Dickinson & Co. BD Diagnostic Systems
- FEI Number
- 1119779
- Product Code
- JSJ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 6, 2013
- Posted
- July 24, 2013
- Terminated
- September 3, 2013
- Address
- 7 Loveton Circle, Sparks, MD, 21152-0999
Description
BD BBL Anaerobe CNA Agar with 5% Sheep Blood // Anaerobe Laked Sheep Blood KV Agar, packaged in cartons of 20 plates, Catalog # 297041, and labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 USA 800-638-8663 www.bd.com/ds*** Anaerobe Laked Sheep Blood KV Agar is used in a qualitative procedure for isolation of gram-negative anaerobic bacilli.
Microbiological identification media may exhibit reduced levels of Vancomycin.
The firm, BD Diagnostic Systems, sent an "URGENT PRODUCT RECALL" letter dated June 2013 to Distributors by fax, US customers via letters by UPS and OUS BD sites via email and teleconference. The letter described the product, problem and actions to be taken. All customers were instructed to discontinue use of the listed product; discard any remaining packages, and complete and return the attached form, whether or not you have any inventory remaining, via Fax to: Attention: Regulatory Compliance, 410-316-4258. BD will issue replacements for the discarded material. It is not necessary to contact BD by telephone to receive replacements. If you have any questions, call 410-316-4000.
Worldwide distribution: US (nationwide) and countries of:Belgium, Canada, Columbia, Mexico, Singapore, and Taiwan.
106 cartons