Product Name: Ingenia 3.0T; Model Numbers: (1) 781342, (2) 781377, (3) 782103;
Recall
- Recall Number
- Z-1797-2025
- Event Number
- 96889
- Firm
- Philips North America Llc
- FEI Number
- 3016618143
- Product Code
- LNH
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- May 13, 2025
- Posted
- May 16, 2025
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2289
Description
Product Name: Ingenia 3.0T; Model Numbers: (1) 781342, (2) 781377, (3) 782103;
Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
On May 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users You may continue to use your system(s) in accordance with the intended use. A. To avoid the potential issue of an incorrect cross reference line in MobiView generated images, use the same voxel size and field of view across all station scans. B. Do not use cross reference line functionality in box mode, all slices mode, or 3D mode. Use the cross reference line functionality in single slice mode. Circulate this notice to all users of this device so that they are aware of the issues and associated hazard/harm. Please retain this Urgent Medical Device Correction letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed. Please complete and return the attached response form to Philips MR promptly and no later than 30 days from receipt of this letter via email to: [email protected]. Completing this form confirms receipt of the Urgent Medical Device Correction letter, understanding of the issues, and required actions to be taken. 5. Actions planned by Philips MR to correct the problem A Philips representative will contact you to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issues (reference FCO78100566, FCO78100584, FCO78100585, FCO78100620, FCO78100621). If you need any further information or support concerning this issue, please contact your local Philips representative. For North America, contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday). This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax. For more information, pl
Worldwide distribution - US Nationwide and the countries of Andorra, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Denmark, Finland, France, French Guiana, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Iceland, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malta, Martinique, Mexico, Moldova, Monaco, Netherlands, New Zealand, Norway, Palestine, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, U.A.E.
193 units (36 US, 157 OUS)