FDA Recall Open, Classified

B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor

Recall: Z-1795-2026 · Initiated March 12, 2026

Recall

Recall Number
Z-1795-2026
Event Number
98561
Firm
Brahms GmbH Neuendorfstr. 25 Hennigsdorf Germany
FEI Number
3003561301
Product Code
QWH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 12, 2026
Posted
April 8, 2026

Description

B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor

Reason

Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.

Action

On 2/27/2026, correction notices were sent to customers informing them: 1. Continue to perform Quality Control testing in accordance with the current Instructions for Use prior to reporting patient results. 2. If Quality Control results fall outside plus-minus 20% of target values follow the troubleshooting procedures described in the User Manual. 3. Forward this letter to those who need to be aware within your organization or to any organization to whom the potentially affected devices have been transferred. Keep this notification on file. For further questions, please contact the firm's customer service on 1-800-232-3342 or e-mail to: [email protected]

Distribution

Worldwide distribution. US states: NC, TX, MN, UT, NY, FL, OH, IL, CT, CA, GA. Other countries: DE, AT, AL, AU, BG, BR, CA, CH, CZ, DK, ET, ES, FR, GB, GR, HK, HU, ID, IL, IT, KW, KZ, LT, MK, ML, MX, NL, NO, PA, PL, PT, RO, SE, SI, TH, TU, TW, UA, ZA

Quantity

5,254