FDA Recall Terminated

CareFusion Safe-T-Centesis Catheter Drainage Tray, Sterile, For Single Use Only, Rx only, Catalog Number PIG1280T, UDI Number 01)10885403108402(17) 210930(10)0001304986

Recall: Z-1795-2020 · Initiated March 26, 2020

Recall

Recall Number
Z-1795-2020
Event Number
85380
Firm
Becton Dickinson & Company
FEI Number
2243072
Product Code
GCB
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
March 26, 2020
Terminated
August 21, 2020
Address
1 Becton Dr, Franklin Lakes, NJ, 07417-1815

Description

CareFusion Safe-T-Centesis Catheter Drainage Tray, Sterile, For Single Use Only, Rx only, Catalog Number PIG1280T, UDI Number 01)10885403108402(17) 210930(10)0001304986

Reason

The 8 Fr catheter was manufactured with the 6.9 Fr needle instead of the 8 Fr needle.

Action

On March 26, 2020, an Urgent Medical Device Recall Notification was sent to consignees stating that Lot #0001304986 of the Safe-T-Centesis 8 Fr Catheter Drainage Trays shipped between June 27, 2019 and September 6, 2019 should be discarded. In addition, they were instructed to complete the Customer Response form regardless of whether or not they have any of the product. Customers with discarded product were instructed to indicate the number of affected products that were discarded on the response form and replacements would be issued to those customers.

Distribution

Domestic distribution to the following states: Alabama, Colorado, Illinois, Massachusetts, Missouri, Oklahoma, Rhode Island, Virginia, and Wisconsin.

Quantity

900