FDA Recall Terminated

Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.

Recall: Z-1795-2013 · Initiated June 5, 2013

Recall

Recall Number
Z-1795-2013
Event Number
65604
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
MQB
Status
Terminated
Root Cause
Device Design
Initiated
June 5, 2013
Posted
July 24, 2013
Terminated
February 24, 2015
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray imaging.

Reason

Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee

Action

On 6/5/13, consignees were notified by letter of a "Field Safety Notice" and received instructions about how to avoid the potential risk of this issue while using the device. Siemens also rolled out a modification of AXIOM Aristos MX that will avoid these potential injuries in the future.

Distribution

Nationwide Distribution.

Quantity

401