FDA Recall Open, Classified

MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.

Recall: Z-1794-2026 · Initiated September 8, 2025

Recall

Recall Number
Z-1794-2026
Event Number
98562
Firm
Beckman Coulter, Inc.
FEI Number
2919016
Product Code
LTT
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
September 8, 2025
Posted
April 8, 2026
Address
2040 Enterprise Blvd, West Sacramento, CA, 95691-3427

Description

MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.

Reason

Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.

Action

On 09/08/2025, the firm sent via mail/email an "URGENT FIELD SAFETY NOTICE" to customers informing them that as a result of an internal investigation Beckman Coulter confirmed that Ciprofloxacin (Cp) antimicrobial agent was incorrectly manufactured on MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08. Customers are instructed to: "Review Ciprofloxacin results generated from Lot 2026-05-08 to identify any abnormal or unexpected trends in both susceptible and resistance patterns, or unexpected change in MIC trends, as compared to historical results for that same species (e.g. a sudden, unexplained increase in susceptible or % susceptible results for a species after implementation of Lot 2026-05-08 versus the previous susceptible or % susceptible results for the same species prior to using Lot 2026-05-08). "Discontinue use of this lot and discard per your laboratory guidelines "The laboratory should retain the inventory of other MicroScan Neg Multidrug Resistant MIC 1 lots as they are not impacted by this issue. Note-Beckman Coulter is no longer distributing MicroScan Neg Multidrug Resistant MIC 1 Lot 2026-05-08 For questions - contact Customer Support Center from website: http://www.beckmancoulter.com

Distribution

International distribution to the country of Poland.

Quantity

3 boxes