FDA Recall Terminated

Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540, 9732622, 9733174, 9733935. As part of kits, the product is either packaged with other instruments or into a holding tray. The Passive Planar Blunt Probe is an optically tracked, reusable instrument used in navigated surgery and is included with Medtronic cranial instrument kits. It is designed for use in patient registration and palpating brain tissue during surgical procedures.

Recall: Z-1794-2015 · Initiated May 5, 2015

Recall

Recall Number
Z-1794-2015
Event Number
71278
Firm
Medtronic Navigation, Inc.
FEI Number
1000517638
Product Code
HAW
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
May 5, 2015
Posted
June 12, 2015
Terminated
July 2, 2015
Address
826 Coal Creek Cir, Louisville, CO, 80027-9710

Description

Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540, 9732622, 9733174, 9733935. As part of kits, the product is either packaged with other instruments or into a holding tray. The Passive Planar Blunt Probe is an optically tracked, reusable instrument used in navigated surgery and is included with Medtronic cranial instrument kits. It is designed for use in patient registration and palpating brain tissue during surgical procedures.

Reason

Passive Planar Blunt Probe (Lot# 150223) was manufactured with a sharp tip rather than a blunt tip.

Action

Medtronic sent an Medical Device Safety Alert letter dated May 5, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken by the user: "Examine your inventory and if any of the affected products are found, please quarantine them for return to Medtronic. "A Medtronic representative will follow-up to arrange for removal and replacement. If you have any questions regarding this action, please contact Medtronic Navigation Technical Services at 1-800-595-9709.

Distribution

US Distribution to the states of : CA, MO, LA and PA., and Internationally to Ireland.

Quantity

18