FDA Recall Open, Classified

Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

Recall: Z-1793-2026 · Initiated February 27, 2026

Recall

Recall Number
Z-1793-2026
Event Number
98374
Firm
Spacelabs Healthcare, Inc.
FEI Number
3010157426
Product Code
DSI
Status
Open, Classified
Root Cause
Component change control
Initiated
February 27, 2026
Posted
April 7, 2026
Address
35301 Se Center St, Snoqualmie, WA, 98065-9216

Description

Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

Reason

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Action

On 2/27/2026, correction notices were emailed to customers who were asked to do the following: 1. Notify all relevant employees within your organization. If products were further distributed, please forward the notice to these sites/customers. 2. Complete and return the acknowledgement form via email to: [email protected] 3. Instruct staff who utilize the cardiac output feature to use the "Manual" mode, or use a non-impacted module in its place (if available) until the impacted module has been corrected. 4. Contact your local Spacelabs Healthcare service and support representative to schedule your device correction as soon as possible. Firm's Support Contact: Phone: 1-800-522-7025 Email: [email protected] Website: https://spacelabshealthcare.com/support/

Distribution

Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.

Quantity

1790