Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Recall
- Recall Number
- Z-1793-2026
- Event Number
- 98374
- Firm
- Spacelabs Healthcare, Inc.
- FEI Number
- 3010157426
- Product Code
- DSI
- Status
- Open, Classified
- Root Cause
- Component change control
- Initiated
- February 27, 2026
- Posted
- April 7, 2026
- Address
- 35301 Se Center St, Snoqualmie, WA, 98065-9216
Description
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
On 2/27/2026, correction notices were emailed to customers who were asked to do the following: 1. Notify all relevant employees within your organization. If products were further distributed, please forward the notice to these sites/customers. 2. Complete and return the acknowledgement form via email to: [email protected] 3. Instruct staff who utilize the cardiac output feature to use the "Manual" mode, or use a non-impacted module in its place (if available) until the impacted module has been corrected. 4. Contact your local Spacelabs Healthcare service and support representative to schedule your device correction as soon as possible. Firm's Support Contact: Phone: 1-800-522-7025 Email: [email protected] Website: https://spacelabshealthcare.com/support/
Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
1790