FDA Recall Open, Classified

TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only

Recall: Z-1792-2024 · Initiated March 19, 2024

Recall

Recall Number
Z-1792-2024
Event Number
94381
Firm
TriMed Inc.
FEI Number
3001236812
Product Code
HWC
Status
Open, Classified
Root Cause
Process control
Initiated
March 19, 2024
Posted
May 10, 2024
Address
27533 Avenue Hopkins, Santa Clarita, CA, 91355-3910

Description

TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only

Reason

Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.

Action

On 4/3/24, TriMed Inc., issued a "Urgent: Medical Device Recall", to affected consignees via E-Mail. TriMed ask consignees to take the following actions: " If you have the affected product, discontinue further use and quarantine products prior to return. " If a successful implantation was completed no further action is necessary. Only standard monitoring of the patient is required after successful implantation. " Please share this notice with all personnel that may utilize the affected products. Please share this notification to any locations where the affected product may have been transferred to. " Conduct a physical count of the affected product in your possession and record the count on attached response form. Once the response form is completed, email to: [email protected]. Please respond even if you have no affected products in your inventory. " Once the response form is completed and emailed, you will receive an RGA (Return Good Authorization form) as well as a link to generate a label to return the devices via email. Please package the devices, print the return label and ship via FedEx.

Distribution

US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.

Quantity

120 units