The Sarns TCM II (system) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. It also freezes water for an ice supply, monitors temperatures in the patient and extracorporeal circuit, and allows gradient rewarming relative to a venous blood temperature. The Sarns" TCM II also features a Cardioplegia System which will supply cooling water for cardioplegia. Device Name / Model Number: TCM II, with Cardioplegia, 110V/60Hz (P/N 164925) TCM II, without Cardioplegia, 110V/60Hz (P/N 164930) TCM II, with Cardioplegia, 220V/50Hz (P/N 164935) TCM II, without Cardioplegia, 220V/50Hz (P/N 164940) TCM (P/N 15747) Catalog Number: 4415, 164925, 164930, 164935, 15747
Recall
- Recall Number
- Z-1791-2021
- Event Number
- 87834
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DWC
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 16, 2021
- Terminated
- November 29, 2023
- Address
- 6200 Jackson Rd, Ann Arbor, MI, 48103-9586
Description
The Sarns TCM II (system) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. It also freezes water for an ice supply, monitors temperatures in the patient and extracorporeal circuit, and allows gradient rewarming relative to a venous blood temperature. The Sarns" TCM II also features a Cardioplegia System which will supply cooling water for cardioplegia. Device Name / Model Number: TCM II, with Cardioplegia, 110V/60Hz (P/N 164925) TCM II, without Cardioplegia, 110V/60Hz (P/N 164930) TCM II, with Cardioplegia, 220V/50Hz (P/N 164935) TCM II, without Cardioplegia, 220V/50Hz (P/N 164940) TCM (P/N 15747) Catalog Number: 4415, 164925, 164930, 164935, 15747
Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.
On 04/30/2021, Terumo issued an Urgent Medical Device Removal notice to customer via letter notifying users to discontinue the use of and dispose of HX2, TCM I and TCM II devices. Customers were instructed to confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. For questions contact Terumo CVS Customer Service: 1-800-521-2818.
Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam
995 devices