FDA Recall Terminated

Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997

Recall: Z-1790-2020 · Initiated March 2, 2020

Recall

Recall Number
Z-1790-2020
Event Number
85282
Firm
CSL Behring GmbH Emil-Von-Behring-Str. 76 Marburg Germany
FEI Number
3003098680
Product Code
NAO
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
March 2, 2020
Terminated
October 5, 2020

Description

Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997

Reason

Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring

Action

Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter PP-20-002.A.US to the 5 affected customers in the U.S. on March 5, 2020 via FedEx overnight delivery. ( Urgent Field Corrective Action (UFCA) PP20-002.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities). Letter states reason for recall, health risk, and action to take by the Customer Please review this letter with your Medical Director. Discontinue use of and discard the kit lots listed in Table 1. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Distribution

IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan

Quantity

297 PCK US; 1247 PCK OUS