FDA Recall Terminated

Jasper Vektor, Class II Correction Appliance, Part No. 610-530R used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.

Recall: Z-1790-2018 · Initiated May 1, 2018

Recall

Recall Number
Z-1790-2018
Event Number
79739
FEI Number
1815540
Product Code
ECO
Status
Terminated
Root Cause
Device Design
Initiated
May 1, 2018
Terminated
April 15, 2020
Address
TP Orthodontics, Inc. 100 Center Plz, La Porte, IN, 46350-9672

Description

Jasper Vektor, Class II Correction Appliance, Part No. 610-530R used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.

Reason

The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient.

Action

On May 3, 2018, the firm sent an Urgent Medical Device Recall letter to customers. The letter advised customers that the firm was voluntarily recalling Generation 1 Jasper Vektor Class 2 Correction Appliances due to receiving 106 customer complaints that the device had broken in the patient's mouth during placement and usage. Customers were instructed to do the following: 1. Locate any of the affected lots identified in the table below and quarantine the product. 2. Return the Generation 1 Jasper Vektor for a replacement Jasper Vektor Generation 2 or refund, following the instructions below: Contact TP Orthodontics, Inc. at 800-338-8856 or [email protected] for a Return Goods Authorization and pre-paid shipping label. Alternately, use the carrier of your choice with available tracking and send to: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 3. Email, mail, or fax the attached response letter to TP Orthodontics, Inc. In addition to sending the letter back, you may call us with any questions or concerns. EMAIL: [email protected] FAX: 219-324-3029 MAIL: TP Orthodontics, Inc. ATTN: Recall Coordinator 100 Center Plaza La Porte, IN 46350 4. If you have any further questions or concerns after reviewing this notice, please contact TPO Customer Care Team 800-388-8856 or [email protected]

Distribution

Distributed to accounts in 20 states: AL, CA, CO, FL, HI, ID, IL, MA, ME, MI, NJ, NV, NY, OH, OR, PA, SC, TX, UT, and VA. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Germany, Spain, Finland, France, Great Britain, Hungary, Ireland, Italy, Japan, Mexico, Netherlands, New Zealand, Portugal, Singapore, and South Africa.

Quantity

7