FDA Recall Terminated

Roche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N). Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.

Recall: Z-1790-2014 · Initiated May 21, 2014

Recall

Recall Number
Z-1790-2014
Event Number
68294
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
JJE
Status
Terminated
Root Cause
Labeling design
Initiated
May 21, 2014
Posted
June 14, 2014
Terminated
June 4, 2015
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N). Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.

Reason

Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.

Action

Roche sent an Urgent Medical Device Correction (UMDC) letter dated May 20, 2014 via UPS Grount (receipt required) to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Follow the instructions in the Precautions Regarding Handling the Aspiration Tube Filter section in this UMDC; 2. Discontinue performing ProCell/Clean Cell aspiration tube filter check as this maintenance item will be performed by your Roche FSR as a part of service maintenance; 3. If you have distributed the products listed within this UMDC to other sites or facilities, please provide them with a copy of the UMDC; and 4. Complete the faxback form enclosed with this UMDC and fax it to 1-888-410-6620. For questions contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.

Distribution

US Nationwide Distribution

Quantity

462