FDA Recall
Open, Classified
Monarch Medical Technologies EndoTool IV, Versions 1.8, Version 1.8.5 and higher, and Version 1.9 and Version 1.10.
Recall: Z-1786-2019
·
Initiated May 3, 2019
Recall
- Recall Number
- Z-1786-2019
- Event Number
- 82828
- Firm
- Monarch Medical Technologies
- FEI Number
- 3009864844
- Product Code
- NDC
- Status
- Open, Classified
- Root Cause
- Software change control
- Initiated
- May 3, 2019
- Address
- 2137 South Blvd, Ste 300, Charlotte, NC, 28203-5189
Description
Monarch Medical Technologies EndoTool IV, Versions 1.8, Version 1.8.5 and higher, and Version 1.9 and Version 1.10.
Reason
Insulin dosing calculations were erroneously high.
Action
The firm notified the consignees by email beginning on 05/03/2019. The notice alerted the consignee of the problem and provided an educational brief and a tip card to provide to the staff to remind them to always use clinical judgement in following EndoTool IV dose recommendations.
Distribution
Nationwide and Dubai
Quantity
83 downloads