FDA Recall Terminated

Accessories for the Captiva Spine FuseLOX Lumbar Cage. The Trial product is a reuseable component of the Captiva Spine FuseLOX Lumbar product. It is not generally sold as an individually packaged product. Trial products are shipped as a component of the FuseLOX Lumbar Caddy which is placed in the FuseLox Lumbar Sterilization tray. The sterilization tray is distributed on consignment. FPT - F is FuseLOX, P is Plif, and T is Trial. Products subject to recall: FuseLOX Lumbar Trials Trial - 7mm x 25mm Part Number: FPT2507 Lot Number 8110001. Trial - 8mm x 25mm Part Number FPT2508 Lot Number 8110002. Trial - 9mm x 25mm Part Number: FPT2509 Lot Number 8110003. Trial - 10mm x 25mm Part Number: FPT2510 Lot Number 8110004. Trial - 11mm x 25mm Part Number: FPT2511 Lot Number 8110005. Trial - 12mm x 25mm Part Number: FPT2512 Lot Number 8110006. Trial - 13mm x 25mm Part Number FPT2513 Lot Number 8110007. Trial - 14mm x 25mm Part Number FPT2514 Lot Number 8110008. The FuseLOX Lumbar Cage implant is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The FuseLOX Lumbar Cage Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine.

Recall: Z-1785-2012 · Initiated April 27, 2012

Recall

Recall Number
Z-1785-2012
Event Number
61925
Firm
Captiva Spine, Inc 967 Alternate A1A Suite1
FEI Number
3006082533
Product Code
MAX
Status
Terminated
Root Cause
Finished device change control
Initiated
April 27, 2012
Posted
June 7, 2012
Terminated
December 24, 2013
Address
Jupiter, FL, 33477

Description

Accessories for the Captiva Spine FuseLOX Lumbar Cage. The Trial product is a reuseable component of the Captiva Spine FuseLOX Lumbar product. It is not generally sold as an individually packaged product. Trial products are shipped as a component of the FuseLOX Lumbar Caddy which is placed in the FuseLox Lumbar Sterilization tray. The sterilization tray is distributed on consignment. FPT - F is FuseLOX, P is Plif, and T is Trial. Products subject to recall: FuseLOX Lumbar Trials Trial - 7mm x 25mm Part Number: FPT2507 Lot Number 8110001. Trial - 8mm x 25mm Part Number FPT2508 Lot Number 8110002. Trial - 9mm x 25mm Part Number: FPT2509 Lot Number 8110003. Trial - 10mm x 25mm Part Number: FPT2510 Lot Number 8110004. Trial - 11mm x 25mm Part Number: FPT2511 Lot Number 8110005. Trial - 12mm x 25mm Part Number: FPT2512 Lot Number 8110006. Trial - 13mm x 25mm Part Number FPT2513 Lot Number 8110007. Trial - 14mm x 25mm Part Number FPT2514 Lot Number 8110008. The FuseLOX Lumbar Cage implant is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The FuseLOX Lumbar Cage Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine.

Reason

Captiva Spine, Inc. initiated a recall of FuseLOX Lumbar Implants, Trials, and Inserters after FDA advised the firm that there is inadequate support for implemented design changes to the cleared device Pivotec Lumbar Interbody Fusion Device.

Action

Captiva Spine sent a Voluntary Correction Notification letter dated May 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate and return affected devices. Customers were also instructed to complete and return a Tracking/Verification Form. For questions customers were instructed to call 877-772-5571. For questions regarding this recall call 561-277-9480.

Distribution

USA (nationwide) including FL,TX, UT, and NJ.

Quantity

54 units