FDA Recall Open, Classified

TASE 500 Imaging systems

Recall: Z-1784-2019 · Initiated January 29, 2019

Recall

Recall Number
Z-1784-2019
Event Number
82988
Firm
COLLINS AEROSPACE
FEI Number
3021214229
Product Code
RET
Status
Open, Classified
Root Cause
Radiation Control for Health and Safety Act
Initiated
January 29, 2019
Address
9 Technology Park Dr, Westford, MA, 01886-3141

Description

TASE 500 Imaging systems

Reason

the Nominal Ocular Hazard Distance (NOHD) associated with a Class lllb alignment laser utilized in its TASE family of imaging systems has been miscalculated.

Action

Collins Aerospace sent a customer notification letter to affected customers,. The letter identified the affected product problem and actions to be taken. For questions contact Customer Support Team at [email protected]

Distribution

US Nationwide Distribution

Quantity

433 in total