FDA Recall Terminated

Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Product Usage: IntelliVue Smart-hopping 1.4 GHz Access Point provides wireless connectivity between IntelliVue devices (listed below) and the Philips IntelliVue Information Center. MX40 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network.

Recall: Z-1780-2013 · Initiated July 8, 2013

Recall

Recall Number
Z-1780-2013
Event Number
65622
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
MHX
Status
Terminated
Root Cause
Software Design Change
Initiated
July 8, 2013
Posted
July 20, 2013
Terminated
May 17, 2017
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Radiohead Access Point 1.4 GHz Radiohead Access Point Firmware versions: C.00.04 & C.00.05 Part Number: 989803171211 Product Usage: IntelliVue Smart-hopping 1.4 GHz Access Point provides wireless connectivity between IntelliVue devices (listed below) and the Philips IntelliVue Information Center. MX40 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network.

Reason

IntelliVue Smart-hopping 1.4 GHz Access Point may experience unexpected signal loss of several minutes impacting wireless connectivity

Action

Philips sent an Urgent Medical Device Correction letter dated July 2013 to affected customers. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer. Philips will perform a firmware upgrade (version C.00..07 for all affected customers at no cost. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377

Distribution

USA Nationwide Distribution

Quantity

20105 units