Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Recall
- Recall Number
- Z-1777-2019
- Event Number
- 82975
- Firm
- Maquet Cardiovascular Us Sales, Llc
- FEI Number
- 3008355164
- Product Code
- MEC
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 20, 2019
- Terminated
- December 15, 2023
- Address
- 45 Barbour Pond Dr, Wayne, NJ, 07470-2094
Description
Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Getinge issued a An Urgent Medical Device Correction to U.S. consignees on 5/20/19 via FedEx 2 day Delivery with Signature Proof of Delivery (SPOD). Letter provides identification of the issue, health risk and action to take: examine your inventory immediately, Until the required correction has been completed, take the following precautions detailed in the User Manual: Do not operate the machine unattended. Always inspect the processed goods after a wash cycle. Contact a Getinge Service person if either of these are noticed: o If the washer starts to sound different than usual during a wash cycle, abort the process and call your service person.o Goods are not cleaned after the process is finished. Getinge Service person will contact the person you listed on the Customer Response Form to schedule service for an update of your device, free of charge. You can also arrange a visit by a Getingetrained or authorized service technician by calling Getinge Customer Support at 18009509912 (select option 1). Customer Service hours: 8:00 AM to 5:00 PM (EST) Monday to Friday
Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine
4 units