FDA Recall Terminated

Conformis iTotal Hip Replacement System: HBS-033-00xx-020101, (HIP-PATIENT SPECIFIC STEM RIGHT)

Recall: Z-1776-2019 · Initiated March 29, 2019

Recall

Recall Number
Z-1776-2019
Event Number
82722
Firm
Conformis, Inc.
FEI Number
3009844603
Product Code
LPH
Status
Terminated
Root Cause
Device Design
Initiated
March 29, 2019
Terminated
April 28, 2020
Address
600 Technology Park Dr, Billerica, MA, 01821-4154

Description

Conformis iTotal Hip Replacement System: HBS-033-00xx-020101, (HIP-PATIENT SPECIFIC STEM RIGHT)

Reason

Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.

Action

Conformis issued on 29 Mar 2019 via e-mails to surgeons with upcoming surgeries the following: 1. For surgeons who have already performed the surgeries (i.e. for patients whose implant surgeries were completed between July 31st, 2018 and March 27th, 2019), a letter sent letter advising them of the issue. Confirmation of receipt will be requested. 2. For surgeons who have not performed the surgeries, but have approved the iView surgical plan on or before March 27, 2019, and Conformis has been notified of confirmed surgery dates or Conformis has provided commitment dates on or before May 17th 2019, a. They will be provided with a marked-up iView surgical plan that provides the true dimensions of the manufactured implant and asked to choose to (i) continue with surgery, (ii) postpone surgery to allow Conformis to manufacture a new implant with a head/neck length matching the original approved iView or (iii) cancel the order. b. Should the physician choose to proceed with the surgery using the manufactured implant with the 3 mm discrepancy, it will be shipped to them after receiving their written approval provided that the surgery is scheduled on or before May 17, 2019. c. Should the physician choose to postpone surgery, a new implant will be manufactured to match the approved iView and provided to the surgeon

Distribution

Nationwide

Quantity

183