FDA Recall Open, Classified

Puritan Bennett 560 Ventilator, PB560, CFN 4096600

Recall: Z-1775-2022 · Initiated August 19, 2022

Recall

Recall Number
Z-1775-2022
Event Number
90766
Firm
Covidien Llc
FEI Number
3002807850
Product Code
QOS
Status
Open, Classified
Root Cause
Process control
Initiated
August 19, 2022
Posted
September 22, 2022
Address
15 Hampshire St, Mansfield, MA, 02048-1113

Description

Puritan Bennett 560 Ventilator, PB560, CFN 4096600

Reason

Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.

Action

Consignees were notified by letter delivered beginning August 19, 2022. Medtronic is advising that continued use of PB560 ventilators is appropriate when the actions listed in the recall letter are taken into consideration, unless this continued use is inconsistent with facility internal policies and procedures. All affected units require turbine replacement. Customers should contact Medtronic Service at 1-800-255-6774 to arrange for required service to replace the affected ventilator turbine component. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.

Distribution

Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.

Quantity

5 US