PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye.
Recall
- Recall Number
- Z-1774-2013
- Event Number
- 65688
- Firm
- Topcon Medical Laser Systems, Inc
- FEI Number
- 2939653
- Product Code
- HQF
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 27, 2012
- Posted
- July 19, 2013
- Terminated
- July 19, 2013
- Address
- 3130 Coronado Dr, Santa Clara, CA, 95054-3205
Description
PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye.
When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.
Initial notification was sent on March 27, 2012 to all TMLS Distribution Partners and Sales Representative. This was followed by a follow up notification on June 4, 2012. A final notice was sent on February 11, 2013 to all those who had not sent documentation that the field action had been implemented.
Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.
7 US, 33 foreign