FDA Recall Open, Classified

XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.

Recall: Z-1773-2026 · Initiated February 18, 2026

Recall

Recall Number
Z-1773-2026
Event Number
98541
Firm
XTANT Medical Holdings, Inc
FEI Number
3005031160
Product Code
MNI
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 18, 2026
Posted
April 3, 2026
Address
664 Cruiser Ln, Belgrade, MT, 59714-9719

Description

XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.

Reason

Units from one lot were mislabeled with an incorrect part number, identifying the device as a shorter screw length than it physically is.

Action

Xtant Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignee on 2/18/2026 via FedEx. The notice explained the reason for the correction, risk to health, and requested the following: 1. Review all devices from Lot 4375-01, including devices in loose stock and devices staged in implant trays or caddies, and verify device identification using the laser-etched markings on each implant to confirm that devices are stored or staged in the appropriate location consistent with their size and configuration. 2. If any device retains non-conforming labeling, relabel the device and discard of the incorrect label and application of the corrected labeling provided by Xtant Medical. 3. No product return is required unless a device cannot be confidently verified or corrected on site; please contact Xtant Medical for further guidance in such cases. 4. Complete Correction Acknowledgement to confirm completion of the above actions. For questions, Rebecca Lennemann, SVP, Quality & Regulatory Affairs, at 1-406-924-5878, Monday through Friday, 8 a.m. to 5 p.m. (Mountain Time).

Distribution

US: OR

Quantity

2 units