FDA Recall
Terminated
Infusion System. EKOS EndoWave Infusion System. EKOS EndoWave Infusion System Drug Delivery Catheter with Ultrasound Core. Labeled as Sterile, Single Use Only. Connector clip is CIC Connector Clip. The infusion system is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytes, into the peripheral vasculature.
Recall: Z-1773-2008
·
Initiated April 1, 2008
Recall
- Recall Number
- Z-1773-2008
- Event Number
- 47785
- Firm
- EKOS Corporation
- FEI Number
- 3001627457
- Product Code
- KRA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 1, 2008
- Posted
- August 20, 2008
- Terminated
- May 13, 2011
- Address
- 11911 N Creek Pkwy S, Bothell, WA, 98011-8809
Description
Infusion System. EKOS EndoWave Infusion System. EKOS EndoWave Infusion System Drug Delivery Catheter with Ultrasound Core. Labeled as Sterile, Single Use Only. Connector clip is CIC Connector Clip. The infusion system is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytes, into the peripheral vasculature.
Reason
Catheter disconnection. There is a potential for the catheter to disconnect from the Catheter Interface Cable (CIC) pod which connects to the Control Unit.
Action
Firm representatives will visit and distribute connector clips and instructions for use to consignees beginning 04/01/2008 at no cost to the consignee.
Distribution
Class II Recall - Nationwide
Quantity
218 units