FDA Recall Open, Classified

MicroScan Neg MIC 3J REF C54814

Recall: Z-1772-2026 · Initiated June 25, 2025

Recall

Recall Number
Z-1772-2026
Event Number
98553
Firm
Beckman Coulter, Inc.
FEI Number
2919016
Product Code
LTT
Status
Open, Classified
Root Cause
Process control
Initiated
June 25, 2025
Posted
April 3, 2026
Address
2040 Enterprise Blvd, West Sacramento, CA, 95691-3427

Description

MicroScan Neg MIC 3J REF C54814

Reason

Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics

Action

On 07/15/2024, firm sent via mail/email an "URGENT MEDICAL DEVICE RECALL" Letter to customer inform them that customers have reported multiple patient samples with skipped wells for Ceftolozane/Tazobactam (C/T) when tested with MicroScan Neg MIC 3J Lot 2025-08-23. The C/T 2/4 dilution well exhibited no growth while growth was observed in the C/T 4/4 well. Customers are instructed to: "Discontinue use of this lot and discard per your laboratory guidelines. "Consider reviewing the C/T antimicrobial susceptibility test results from this lot if Enterobacterales reported as resistant and P. aeruginosa reported as intermediate or if an increasing trend in resistance has been observed for C/T in your laboratory. "The laboratory should retain the inventory of other MicroScan Neg MIC 3J lots as they are not impacted by this issue. any questions regarding this notice, please contact us: "From our website: http://www.beckmancoulter.com "By phone: please contact: Customer call center TEL :0120-566-730

Distribution

International distribution in the country of Japan.

Quantity

688 boxes (20 panels per box)